Advanced Battery Technology
EMC Notebook

Design Impact of IEC 60601, Second Edition

By William D. Kimmel, P.E.
and Daryl D. Gerke, P.E.

Kimmel Gerke Associates, Ltd.

We try to stay away from any articles on EMC regulations – our specialty is EMC design and troubleshooting and there are better sources of information on regulations . But the emergence of new requirements inevitably impacts the design people, and that brings us into the picture. A case in point is the adoption of IEC 60601-1-2, second edition.

Medical device EMC (and safety) is covered by IEC 60601-1-2, which is recognized by the FDA (in the U.S.), in the European Union, and many other countries (under various names). The original edition, dated 1993, will soon be superseded by the second edition, dated 2001. The transition time, where use of the first edition was still permitted, comes to an end in November.

Figure 1. Protecting isolated circuits from RFI



The first edition had numerous deficiencies, including references to outdated standards and unclear pass/fail criteria. The second edition is greatly expanded, answering a lot of questions raised by the first edition (and, of course, raising more questions). We’re not going to read the requirements to you, but we will highlight a few of the changes that significantly affect designers.

Medical electronics encompasses a wide range of equipment and environments. There are cases where IEC 60601 will not be adequate, and other cases where the specified levels are not achievable. It will take a specialist to make a determination of what the requirements are for your particular equipment and how to design for them, but here are some basic guidelines.

Essential Performance

The biggest single aspect is the incorporation of “essential performance,” a definition of failure. In the first edition, the interpretation was the equipment could “fail safe” – as long as the equipment did not become unsafe, it was deemed to have passed the test. This, of course, resulted in interrupted operation, and was generally unsatisfactory. The second edition specifies that the equipment must meet its intended function, but the requirements are not very well worded. We like the viewpoint given by the FDA(1), which is pretty explicit, and is given below:

The following degradations are not allowed:


Component Failures

Changes in programmable parameters

Reset to factor defaults (manufacturer’s presets); change of operating mode

False alarms

Cessation of any intended operation, even if accompanied by an alarm

Initiation of any unintended operation, including unintended or uncontrolled motion, even if accompanied by an alarm

Noise on a waveform in which the noise is indistinguishable from physiologically-produced signals or the noise interferes with interpretation of physiologically-produced signals

Artifact or distortion in an image in which the artifact is indistinguishable from physiologically-produced signals or the distortion interferes with interpretation of physiologically-produced signals

Error of a displayed numerical value sufficiently large to affect diagnosis, therapy or treatment

Failure of an automatic diagnosis or treatment equipment and/or systems to diagnose or treat, even if accompanied by an alarm

We think that pretty well covers it – it won’t give the lawyers much room to wiggle, will it?


ESD

The ESD requirements have been increased from 3kV to 6kV (contact), which will have a definite impact on design – lots of equipment that passed the 3kV will crash at 6kV.


RFI

The RFI immunity level is increased from 3V/m to 10V/m and conducted immunity is increased from 3V to 10v (in the ISM bands, approximately 6.7, 13.5, 27 and 41MHz) for life supporting equipment. At 3V/m, the impact was largely on the sensitive analog circuits, but, with the increase, we expect to see more problems showing up with digital electronics, as well.

There are a number of new requirements, including magnetic fields, flicker and harmonics, voltage dips and interruptions. In the EU, these requirements are generally applied over all electronics, not just medical electronics. The requirements have a moderate effect on the design, depending on the equipment. Here are some observations:

Harmonics on power lines arise from non-linear loads in your power supply and, of course, do not apply to battery-powered devices. Voltage dips and interruptions also involve the power supply.

Flicker only occurs with significant load changes in the equipment – not an issue with most electronic equipment. Magnetic fields are generally of interest only with a few low impedance devices, notably the CRT.

Systems Aspect

The second edition includes a discussion of system aspects, as well as equipment aspects. This is in recognition of the fact that equipment sensitivities and needs vary widely. Some equipment will need to be operated in a shielded room, either because it is too sensitive to achieve the immunity levels spelled out, or because it is too noisy. Other equipment will not need such protection, but will need to be operated some distance away from radio sources – so a calculation of recommended separation is required.

How is Design Affected?

As this is a design column, we would like to provide a glimpse as to how these changes impact the design.

You first need to look at how “essential performance” impacts your design philosophy. If you have already made the interpretation that the equipment needs to work during interference, you won’t find a difference. Otherwise, you will be confronted with some significant changes.

Having decided that, your biggest problems are likely to be RFI and ESD.

RFI

Regardless of what level of RFI you need to meet, the big problem will remain the sensitive analog input devices. These amplifiers, especially the first few sensitive ones, may be sensitive to signals in the millivolt or even the microvolt range – even 3V/m may well be too much for your particular application.

Figure 2. ESD path through seem in plastic.


Add to this the case where the equipment is to be patient-connected. Physiological signal levels are quite low, some down in the low microvolt range, which means that you will have sensitive input amplifiers connected to patient cables that can’t be effectively shielded – so filtering is the only real option. As we are faced with isolation requirements, the filters need to be referenced to isolated ground (Figure 1) – then you need to boost the signal to the point where you can bridge the isolation barrier.

ESD

ESD poses a problem, especially with portable plastic enclosures. If you have a plastic enclosure, ESD can only enter at seams and openings – there you look for possible discharge paths to a metal element inside the enclosure (Figure 2). All too often we find a box has been metallized to contain RF emissions or immunity, only to find new ESD points at the seams (Figure 2 lower). We find the best way to work ESD is to recess the metal members or protect them with a dielectric. If you can avoid a direct discharge path, your ESD problem is reduced to an indirect coupling path, a much lesser problem. If you do find you need a metallized enclosure and can’t isolate the contact points, you will need to do a very good job of shielding.

Summary

The second edition of IEC 60601-1-2 incorporates some significant changes – a major improvement, in our opinion. If you haven’t been minding the store, you are running out of time – your new designs will have to be improved.

The big change is the interpretation of pass/fail – you’ll have to make the equipment work. RFI always poses a problem to sensitive analog input devices – this problem will increase when systems are exposed to the higher levels of RFI and ESD test levels will also increase, posing a problem for perhaps the first time.

References

1) The Draft Second Edition of IEC 60601-1-2: A further Update, Jeffrey L. Silberberg, IEEE International Symposium on EMC, August, 1999.

2) IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Safety – Collateral Standard - Electromagnetic Compatibility – Requirements for Tests.

N E X T -->
<-- B A C K

 

  <-- B A C K T O P N E X T -->